Question 1

In which of the following patients would an end-tidal CO2most likely produce values that are inconsistent with PaCO2?
  • Question 2

    A helium dilution test has just been performed on a patient. The following results are obtained:
    FRC 5.0 L
    VC 4.0 L
    ERV 1.5 L
    TLC was calculated to be 6.0 L by plethysmography. From this information, a pulmonary function technologist should conclude that the patient
  • Question 3

    A pulmonary function technologist is performing quality control on a nebulizer used in the 5breath dosimeter bronchial challenge. The target output of the device is 0.09 mL, plus or minus 10%. After 10 actuations, the nebulizer output was 75uL with a 2.0 mL initial saline dose in the nebulizer. The technologist should
  • Question 4

    During daily quality control procedures on an infrared CO2analyzer, a pulmonary function technologist is unable to adjust the gain to the calibration gas concentration. Which of the following is the most likely explanation?
  • Question 5

    When performing quality control in a body plethysmograph using a 5-L isothermal bottle, the VTG at shutter closure are as follows:

    A pulmonary function technologist should