Question 1
Which of the following problems may be identified by using an isothermal lung analog to perform quality control on a body plethysmograph?
1. Improperly calibrated mouth pressure transducer
2. Obstructed or perforated pneumotachometer
3. Increase in mechanical resistance
4. Malfunctioning box pressure transducer
1. Improperly calibrated mouth pressure transducer
2. Obstructed or perforated pneumotachometer
3. Increase in mechanical resistance
4. Malfunctioning box pressure transducer
Question 2
The following test results are available for a 35-year-old subject who is applying for disability:
These findings are consistent with
These findings are consistent with
Question 3
A pulmonary function technologist is reviewing an asthma action plan. The patient takes a low-dose inhaled corticosteroid and a long-acting beta2-agonist twice daily. The patient's personal best peak flow is 600 L/min.
Which of the following instructions should be given to the patient when the peak flow drops to 500 L/min?
Which of the following instructions should be given to the patient when the peak flow drops to 500 L/min?
Question 4
The following data are obtained after an exercise (stress) test for exercise-induced asthma:

Based on these data, the post-exercise FEVi represents a decrease of approximately

Based on these data, the post-exercise FEVi represents a decrease of approximately
Question 5
An isothermal lung analog has the following specifications:
5-L glass bottle 2000 g of steel wool Density of steel wool 5 g/cm3
Volume of connectors and bulb 100 mi-Repeated FRC measurements in the plethysmograph result in a volume of 5.5 L.
Which of the following should a pulmonary function technologist do?
5-L glass bottle 2000 g of steel wool Density of steel wool 5 g/cm3
Volume of connectors and bulb 100 mi-Repeated FRC measurements in the plethysmograph result in a volume of 5.5 L.
Which of the following should a pulmonary function technologist do?